ABB Multiwave ultraviolet process photometer PUV3402

The ABB Multiwave ultraviolet process photometer PUV3402 is designed for the UV (ultraviolet) and VIS (visible) spectra regions. The ABB Multiwave photometers provide continuous measurements in complex streams which allow for fast reactions to process changes. The fast analysis provides optimum control of a process and improves its safe operation. Measurements can be made in liquid or vapor streams.

ABB Multiwave ultraviolet process photometer PUV3402

ABB Multiwave ultraviolet process photometer photo

The PUV3402 multiwave process photometer provides continuous measurements of vapor or liquid samples, in simple or complex process streams, for process control, product quality assurance, safety, catalyst protection and area monitoring.

The filter wheel of the photometer utilizes up to eight wavelengths, increasing the number of measurement solutions. Because it uses multiple wavelengths, the photometer can compensate for many types of interference and handle multiple component applications.

The PUV3402 process photometers are designed for the UV (Ultraviolet) and VIS (visible) spectra regions.  The PUV3402 provides the following:

  • Continuous measurements
  • Multiple component samples
  • Measure vapor or liquid samples
  • Operates in the UV and VIS
  • Multiple interference compensation capability
  • VistaNET and VistaSTAR Connectivity

Applications for the ABB Multiwave ultraviolet process photometer

Multiwave photometer solutions

The ABB PUV3402 process photometers leading applications include:

Visible

Color Measurements (ASTM, APHA, Saybolt Color Units)
With the Spectral Regions: 400 – 800 nm

Ultraviolet

%T @Multiple Wavelengths in Glycol Process
Chlorine in Phosgene
With the Spectral Regions: 200 – 400 nm

Specifications of the ABB Multiwave ultraviolet process photometer

Performance Specifications

Precision: ± 1 % of full scale
Noise: ± 1 % of full scale at 0.02 absorption units
± 0.5 % of full scale at 0.20 absorption units
Linearity: Standard ± 2 % of full scale
Zero Drift: ± 1.0% of full scale per day for IR;
   

Operating Specifications

Ambient Temperature Range: 32° to 113°F (0–45°C)
Sample Pressure: 0–500 PSIG (0–34 bar)
Voltage Input Variation: 10 % fluctuation without causing an output variation of 0.05 % of full scale
Electrical Power to Analyzer: 100/115/220/230 VAC, 45–66 HZ; 150W, Maximum Power Consumption; 600W With Electric Heat
   

General Installation

  Protect the instrument from direct sunlight and rain at operating temperature between 32°–113° F (0–45°C)
   

Dimensions

Weight: approximately 80 lbs (36.28 kg)
Height: 13.5 inches (342.9 mm)
Depth: 10.5 inches (266.7 mm)
Width: 10 inches (254 mm) per module
Overall Length: (based on cell pathlength)
Minimum: 26.8 inches (681.4 mm) for 0.5 to 16 mm cell
Maximum: 65.5 inches (1665 mm) for 1 meter cell
   

Area Classification

Standard: NEC Class 1, Division 2, Group B, C, D
Enclosure Purge: NEC Class 1, Group B, C, D, Division 1, with “Y”
Purge CENELEC Zone 1, IIB + H2

Contact Our specialist today

For personalized assistance and further inquiries, please contact our specialist at ASaP. We are here to help and provide the support you need.

Contact our specialist directly

You can contact us is you have a question about when it suits you. See our specialist’s contact information below.

Mike Beer - ASaP - Address and contact persons

🔧

📨

📞

🔗

Let our specialist contact you

Let us know how we can reach you to discuss . We will only use this information for contacting you.

Contact form